An official Health Canada notice concerns Benchmark ultra stainer module. The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury. Recall start date: Jan 18, 2022 Published 2022-02-08.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63904
Official notice
2022-02-08
The hazard
The hazard
The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury. Recall start date: Jan 18, 2022
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Ventana Medical, Benchmark Ultra Stainer Module.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Ventana he 600 system

Pas staining kit

Ventana® anti-cd10 (sp67) rabbit monoclonal primary antibody

Solar™ gi
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 8 Feb 2022. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.