Medium riskHealth Canada

An official Health Canada notice concerns Benchmark ultra stainer module. The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury. Recall start date: Jan 18, 2022 Published 2022-02-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63904

Official notice

2022-02-08

The hazard

The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury. Recall start date: Jan 18, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63904

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Benchmark Ultra Stainer Module, Lot or Serial Number - More than 10 numbers, contact manufacturer., Model or catalog number - N750-BMKU-FS

Risk / hazard

The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury. Recall start date: Jan 18, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ventana Medical, Benchmark Ultra Stainer Module.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

63904

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Ventana he 600 system

Medium riskHealth Canada18 Jul 2022
CA
No image published for this recall

Pas staining kit

Medium riskHealth Canada18 Jul 2022
CA
No image published for this recall

Ventana® anti-cd10 (sp67) rabbit monoclonal primary antibody

Medium riskHealth Canada25 Jul 2024
CA
No image published for this recall

Solar™ gi

Critical riskHealth Canada16 Jul 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 8 Feb 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.