Beckman coulter au immunochemistry system - bicarbonate (co2) assay

Official notice

13 Apr 2023

Last checked

13 Apr 2023

Medium riskHealth Canada

An official Health Canada notice concerns Beckman coulter au immunochemistry system - bicarbonate (co2) assay. Beckman Coulter has confirmed a premature decrease in calibration ODs for several lots of bicarbonate OSR6X37. It is possible that all bicarbonate lots may also be impacted by this issue, therefore all lots are subject to this field action. The abnormal decrease in calibrator ODs may result in calibration failures and/or OC failures. The risk of generating falsely elevated, clinically impactful patient test results is remote. Recall start date: April 3, 2023 Published 2023-04-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73500

Official notice

2023-04-13

The hazard

Beckman Coulter has confirmed a premature decrease in calibration ODs for several lots of bicarbonate OSR6X37. It is possible that all bicarbonate lots may also be impacted by this issue, therefore all lots are subject to this field action. The abnormal decrease in calibrator ODs may result in calibration failures and/or OC failures. The risk of generating falsely elevated, clinically impactful patient test results is remote. Recall start date: April 3, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73500

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Beckman Coulter Au Immunochemistry System - Bicarbonate (CO2) Assay, Lot or serial number - All lots., Model or catalog number - OSR6237 OSR6137

Product 2

Affected products - Beckman Coulter Au2700 Immunochemistry System - Bicarbonate(CO2) Assay, Lot or serial number - All lots., Model or catalog number - OSR6237 OSR6137

Product 3

Affected products - Beckman Coulter Au5400 Immunochemistry System - Bicarbonate (CO2)Assay, Lot or serial number - All lots., Model or catalog number - OSR6237 OSR6137

Risk / hazard

Beckman Coulter has confirmed a premature decrease in calibration ODs for several lots of bicarbonate OSR6X37. It is possible that all bicarbonate lots may also be impacted by this issue, therefore all lots are subject to this field action. The abnormal decrease in calibrator ODs may result in calibration failures and/or OC failures. The risk of generating falsely elevated, clinically impactful patient test results is remote. Recall start date: April 3, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Beckman Coulter, Beckman Coulter Au Immunochemistry System - Bicarbonate Assay.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

73500

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Apr 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.