No image published for this recall

No image published for this recall

Bd pcr cartridges 24

Official notice

2 Jan 2025

Last checked

2 Jan 2025

Medium riskHealth Canada

An official Health Canada notice concerns Bd pcr cartridges 24. BD has identified through customer complaint investigations, that signal drift has been observed in specific lots of BD PCR cartridges. BD has identified the following impacts: • for the BD Max™ RVP assay (catalog # 445215 & 445373) signal drift could cause false positive results for Influenza A. • for specific BD Max™ molecular assays including enteric bacterial panel (EBP, catalog #442963), enteric parasite panel (EPP, catalog #442960), StaphSR (catalog #443419), MDR-TB (catalog #443878), CPO (catalog # 445262) signal drift could potentially cause an increase in unresolved results (UNR). • for the BD COR™ VP (catalog # 443976) signal drift could potentially cause an increase in unresolved results (UNR). • for third-party assays used with the open system reagent kits (BD MAX™ ExK™ DNA- 1, DNA-2, DNA-3, DNA MMK, DNA MMK (SPC), TNA MMK, TNA MMK (sSPCpc), TNA-2 and TNA-3) signal drift could potentially cause an increase in unresolved results (UNR) or unexpected results. Recall start date: December 19, 2024 Published 2025-01-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76789

Official notice

2025-01-02

The hazard

BD has identified through customer complaint investigations, that signal drift has been observed in specific lots of BD PCR cartridges. BD has identified the following impacts: • for the BD Max™ RVP assay (catalog # 445215 & 445373) signal drift could cause false positive results for Influenza A. • for specific BD Max™ molecular assays including enteric bacterial panel (EBP, catalog #442963), enteric parasite panel (EPP, catalog #442960), StaphSR (catalog #443419), MDR-TB (catalog #443878), CPO (catalog # 445262) signal drift could potentially cause an increase in unresolved results (UNR). • for the BD COR™ VP (catalog # 443976) signal drift could potentially cause an increase in unresolved results (UNR). • for third-party assays used with the open system reagent kits (BD MAX™ ExK™ DNA- 1, DNA-2, DNA-3, DNA MMK, DNA MMK (SPC), TNA MMK, TNA MMK (sSPCpc), TNA-2 and TNA-3) signal drift could potentially cause an increase in unresolved results (UNR) or unexpected results. Recall start date: December 19, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76789

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - BD PCR Cartridges 24, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 437519

Risk / hazard

BD has identified through customer complaint investigations, that signal drift has been observed in specific lots of BD PCR cartridges. BD has identified the following impacts: • for the BD Max™ RVP assay (catalog # 445215 & 445373) signal drift could cause false positive results for Influenza A. • for specific BD Max™ molecular assays including enteric bacterial panel (EBP, catalog #442963), enteric parasite panel (EPP, catalog #442960), StaphSR (catalog #443419), MDR-TB (catalog #443878), CPO (catalog # 445262) signal drift could potentially cause an increase in unresolved results (UNR). • for the BD COR™ VP (catalog # 443976) signal drift could potentially cause an increase in unresolved results (UNR). • for third-party assays used with the open system reagent kits (BD MAX™ ExK™ DNA- 1, DNA-2, DNA-3, DNA MMK, DNA MMK (SPC), TNA MMK, TNA MMK (sSPCpc), TNA-2 and TNA-3) signal drift could potentially cause an increase in unresolved results (UNR) or unexpected results. Recall start date: December 19, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Becton Dickinson And [bd], Bd Pcr Cartridges 24.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

76789

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 2 Jan 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.