Bd alaris gateway workstation

Official notice

21 Jul 2023

Last checked

21 Jul 2023

Medium riskHealth Canada

An official Health Canada notice concerns Bd alaris gateway workstation. BD is issuing a recall because product testing demonstrated that the BD Alaris™ Gateway Workstation can no longer claim protection against fluid ingress according to international standard, ip22/ipx1. Therefore, BD is providing a notice to affected customers to re-enforce the use of the existing directions for use (DFU) and technical service manual (TSM) to minimize potential fluid ingress. During the evaluation of the AGW, the unit was subject to water as per iec 60529 test requirements for ipx2. As part of clause 11.6.5 from iec 60601-1, after the tests of iec 60529, the device is to show no signs of bridging of insulation that is likely to result in the loss of basic safety or essential performance in normal condition or in combination with a single fault condition. The single fault condition identified prior to starting the test was a failure of the pump detection circuit, leading to voltage being on the iec outlet on the tile, when no pump is present. This single fault condition in combination with water being present on the iec c13 outlet, where water could conduct a path from the live terminal to an operator, resulted in ul deeming the test a failure. The DFU contains a claim for ip22 protection. Recall start date: July 7, 2023 Published 2023-07-21.

What to do now

Review notice & distribute to users, follow instructions provided, complete/ return response form.

Recall number

73969

Official notice

2023-07-21

The hazard

BD is issuing a recall because product testing demonstrated that the BD Alaris™ Gateway Workstation can no longer claim protection against fluid ingress according to international standard, ip22/ipx1. Therefore, BD is providing a notice to affected customers to re-enforce the use of the existing directions for use (DFU) and technical service manual (TSM) to minimize potential fluid ingress. During the evaluation of the AGW, the unit was subject to water as per iec 60529 test requirements for ipx2. As part of clause 11.6.5 from iec 60601-1, after the tests of iec 60529, the device is to show no signs of bridging of insulation that is likely to result in the loss of basic safety or essential performance in normal condition or in combination with a single fault condition. The single fault condition identified prior to starting the test was a failure of the pump detection circuit, leading to voltage being on the iec outlet on the tile, when no pump is present. This single fault condition in combination with water being present on the iec c13 outlet, where water could conduct a path from the live terminal to an operator, resulted in ul deeming the test a failure. The DFU contains a claim for ip22 protection. Recall start date: July 7, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73969

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - BD Alaris Gateway Workstation, Lot or serial number - All lots., Model or catalogue number - 80203UNS02-30 80203UNS02-34 80203UNS02-54 80203UNS02-74 80203UNS03-34 80203UNS03-52 80203UNS03-53 80300UNS02-53 80300UNS02-72 80203UNS01-32 80203UNS01-53 80203UNS02-50 80203UNS02-72 80203UNS02-73 80300UNS02-32 80203UNS01-235 80203UNS01-72 80203UNS02-235 80203UNS03-30 80203UNS03-32 80300UNS02-235 80203UNS01-34 80203UNS01-50 80203UNS01-92 80203UNS02-33 80203UNS03-54 80300UNS02-34

Product 2

Affected products - BD Alaris Gateway Workstation, Lot or serial number - All lots., Model or catalogue number - 80203UNS01-30 80203UNS01-33 80203UNS01-73 80203UNS02-52 80203UNS02-92 80203UNS03-74 80203UNS01-52 80203UNS01-70 80203UNS02-035 80203UNS02-53 80203UNS03-72 80300UNS02-73 80203UNS01-54 80203UNS03-33 80203UNS03-50 80300UNS02-50 80300UNS02-54 80300UNS02-70 80300UNS02-74 80300UNS02-92 80203UNS01-74 80203UNS02-32 80203UNS02-70 80203UNS03-70 80203UNS03-73 80300UNS02-035 80300UNS02-33 80300UNS02-52

Risk / hazard

BD is issuing a recall because product testing demonstrated that the BD Alaris™ Gateway Workstation can no longer claim protection against fluid ingress according to international standard, ip22/ipx1. Therefore, BD is providing a notice to affected customers to re-enforce the use of the existing directions for use (DFU) and technical service manual (TSM) to minimize potential fluid ingress. During the evaluation of the AGW, the unit was subject to water as per iec 60529 test requirements for ipx2. As part of clause 11.6.5 from iec 60601-1, after the tests of iec 60529, the device is to show no signs of bridging of insulation that is likely to result in the loss of basic safety or essential performance in normal condition or in combination with a single fault condition. The single fault condition identified prior to starting the test was a failure of the pump detection circuit, leading to voltage being on the iec outlet on the tile, when no pump is present. This single fault condition in combination with water being present on the iec c13 outlet, where water could conduct a path from the live terminal to an operator, resulted in ul deeming the test a failure. The DFU contains a claim for ip22 protection. Recall start date: July 7, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Bd Switzerland Sarl, Bd Alaris Gateway Workstation.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

73969

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 21 Jul 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.