Battery mobile x-ray systems soltus

Official notice

14 Dec 2022

Last checked

14 Dec 2022

Medium riskHealth Canada

An official Health Canada notice concerns Battery mobile x-ray systems soltus. There have been incidents of erratic movement involving Soltus mobile x-ray units. Shis problem is caused by a software problem: the backwards high-speed movement is produced due to a bug in the movement control loop while the bumper is activated, the user pushes "manually" the mobile, and then at a certain moment the bumper is mechanically released. In this situation, the reference values for the movement control loop were not correct, and the unit moved backwards and rotates in high speed. Recall start date: November 18, 2022 Published 2022-12-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72336

Official notice

2022-12-14

The hazard

There have been incidents of erratic movement involving Soltus mobile x-ray units. Shis problem is caused by a software problem: the backwards high-speed movement is produced due to a bug in the movement control loop while the bumper is activated, the user pushes "manually" the mobile, and then at a certain moment the bumper is mechanically released. In this situation, the reference values for the movement control loop were not correct, and the unit moved backwards and rotates in high speed. Recall start date: November 18, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72336

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Battery Mobile X-Ray System Soltus 450, Lot or serial number - SLS000191 SLS000192 SLS000193, Model or catalogue number - 10451

Product 2

Affected products - Battery Mobile X-Ray System Soltus 500, Lot or serial number - SLS000131 SLS000130, Model or catalogue number - 10501

Risk / hazard

There have been incidents of erratic movement involving Soltus mobile x-ray units. Shis problem is caused by a software problem: the backwards high-speed movement is produced due to a bug in the movement control loop while the bumper is activated, the user pushes "manually" the mobile, and then at a certain moment the bumper is mechanically released. In this situation, the reference values for the movement control loop were not correct, and the unit moved backwards and rotates in high speed. Recall start date: November 18, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Sociedad Espanola De Electromedicina Y Calidad S.a, Battery Mobile X-ray Soltus.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

72336

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 Dec 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.