No image published for this recall
Azurion and allura systems
Official notice
30 Dec 2024
Last checked
30 Dec 2024
An official Health Canada notice concerns Azurion and allura systems. Philips has identified that due to the forces applied during the movement of the c-arc of the Azurion and Allura systems, the bolts supporting the FlexMove carriage may become loose and/or break, and cracks may appear in the FlexMove carriage. If FlexMove rail fixation bolts become loose or broken, or cracks appear in the carriage, the following issues may occur: • transversal movements of the c-arc stop due to false collision detections because of additional friction. • manual movement of the FlexMove carriage not possible due to added friction. • abnormal noise during transversal movement of the c-arc. • unstable c-arc suspension. • fall of the c-arc assembly (1,500 kg), if all bolts in the x-axis would break/get loose. • drop of the c-arc assembly (up to 10cm if c- arc is at one side of the rail, up to 5 cm if c-arc is in the center of the rails, and 1.5 cm if the c-arc is in the anterior/posterior position), if all bolts in the y-axis would break/get loose. Recall start date: December 12, 2024 Published 2024-12-30.
What to do now
Contact the manufacturer if you require additional information.
Recall number
76771
Official notice
2024-12-30
The hazard
The hazard
Philips has identified that due to the forces applied during the movement of the c-arc of the Azurion and Allura systems, the bolts supporting the FlexMove carriage may become loose and/or break, and cracks may appear in the FlexMove carriage. If FlexMove rail fixation bolts become loose or broken, or cracks appear in the carriage, the following issues may occur: • transversal movements of the c-arc stop due to false collision detections because of additional friction. • manual movement of the FlexMove carriage not possible due to added friction. • abnormal noise during transversal movement of the c-arc. • unstable c-arc suspension. • fall of the c-arc assembly (1,500 kg), if all bolts in the x-axis would break/get loose. • drop of the c-arc assembly (up to 10cm if c- arc is at one side of the rail, up to 5 cm if c-arc is in the center of the rails, and 1.5 cm if the c-arc is in the anterior/posterior position), if all bolts in the y-axis would break/get loose. Recall start date: December 12, 2024
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips, Azurion And Allura.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 30 Dec 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.