An official Health Canada notice concerns Aquilex fluid control system. WOM has received complaints from customers in the United States that the display of inflow volume on the Aquilex Fluid Control System can reach its limit of 30,000 ml during long procedures. In these cases, the display does not indicate the actual inflow volume. In cases where an unusually high amount of fluid is used for distension of the patient's uterus, the calculation of the inflow and outflow volumes reaches its limit and the pump control unit display will freeze at the maximum value. There is a risk of distension fluid reaching the circulatory system of the patient's soft tissue. Recall start date: July 17, 2023 Published 2023-08-15.
What to do now
Contact the manufacture if you require additional information.
Recall number
74123
Official notice
2023-08-15
The hazard
The hazard
WOM has received complaints from customers in the United States that the display of inflow volume on the Aquilex Fluid Control System can reach its limit of 30,000 ml during long procedures. In these cases, the display does not indicate the actual inflow volume. In cases where an unusually high amount of fluid is used for distension of the patient's uterus, the calculation of the inflow and outflow volumes reaches its limit and the pump control unit display will freeze at the maximum value. There is a risk of distension fluid reaching the circulatory system of the patient's soft tissue. Recall start date: July 17, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: W.o.m World Of Medicine, Aquilex Fluid Control System.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 15 Aug 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.