Allura xper systems

Official notice

24 Sept 2025

Last checked

24 Sept 2025

Medium riskHealth Canada

An official Health Canada notice concerns Allura xper systems. Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted. Published 2025-09-24.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78125

Official notice

2025-09-24

The hazard

Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78125

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Allura Xper FD20 (Product Of AlluraClarity Family), Lot or serial number - Not applicable., Model or catalogue number - 722028

Product 2

Affected products - Allura Xper FD 20 Or Table (Product Of AlluraClarity Family), Lot or serial number - Not applicable., Model or catalogue number - 722035

Product 3

Affected products - Allura Xper FD10/10 (Product Of AlluraClarity Family), Lot or serial number - Not applicable., Model or catalogue number - 722027

Product 4

Affected products - Allura Xper FD10 (Product Of AlluraClarity Family), Lot or serial number - Not applicable., Model or catalogue number - 722026

Product 5

Affected products - Allura Xper FD20/20 (Product Of AlluraClarity Family), Lot or serial number - Not applicable., Model or catalogue number - 722038

Product 6

Affected products - Allura Xper FD10, Lot or serial number - Not applicable., Model or catalogue number - 001443

Product 7

Affected products - Allura Xper FD20/15, Lot or serial number - Not applicable., Model or catalogue number - 722058

Risk / hazard

Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips, Allura Xper.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

78125

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 24 Sept 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.