Critical riskHealth Canada

An official Health Canada notice concerns Allura xper systems. Philips has identified that a component (air baffle) may be potentially missing in some Nehalem FlexVision-2 PCs that are part of the Allura XPER series. The air baffle directs part of the airflow inside the PC. When it is not installed, the temperature of the frame grabber card of the FlexVision PC-2 may be outside the specified temperature range, which may lead to an increased probability of failure of the frame grabber card. If this occurs, no image will be shown in the FlexVision monitor of the exam room. Recall start date: August 1, 2025 Published 2025-08-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77920

Official notice

2025-08-15

The hazard

Philips has identified that a component (air baffle) may be potentially missing in some Nehalem FlexVision-2 PCs that are part of the Allura XPER series. The air baffle directs part of the airflow inside the PC. When it is not installed, the temperature of the frame grabber card of the FlexVision PC-2 may be outside the specified temperature range, which may lead to an increased probability of failure of the frame grabber card. If this occurs, no image will be shown in the FlexVision monitor of the exam room. Recall start date: August 1, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77920

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Allura Xper FD20/20 (Product Of Alluraclarity Family), Lot or serial number - Not applicable., Model or catalog number - 722038

Product 2

Affected products - Allura Xper FD20 OR Table (Product Of Alluraclarity Family), Lot or serial number - Not applicable., Model or catalog number - 722058

Product 3

Affected products - Allura Xper FD10, Lot or serial number - 608, Model or catalog number - 001443

Product 4

Affected products - Allura Xper FD10/10 (Product Of Alluraclarity Family), Lot or serial number - Not applicable., Model or catalog number - 722027

Product 5

Affected products - Allura Xper FD20/10 (Product Of Alluraclarity Family), Lot or serial number - Not applicable., Model or catalog number - 722029

Product 6

Affected products - Allura Xper FD10 (Product Of Alluraclarity Family), Lot or serial number - 1115, Model or catalog number - 722026

Product 7

Affected products - Allura Xper FD20 (Product Of Alluraclarity Family), Lot or serial number - 1094, Model or catalog number - 722028

Product 8

Affected products - Allura Xper FD20 (Product Of Alluraclarity Family), Lot or serial number - 1681, Model or catalog number - 722028

Product 9

Affected products - Allura Xper FD20 (Product Of Alluraclarity Family), Lot or serial number - 2282, Model or catalog number - 722028

Risk / hazard

Philips has identified that a component (air baffle) may be potentially missing in some Nehalem FlexVision-2 PCs that are part of the Allura XPER series. The air baffle directs part of the airflow inside the PC. When it is not installed, the temperature of the frame grabber card of the FlexVision PC-2 may be outside the specified temperature range, which may lead to an increased probability of failure of the frame grabber card. If this occurs, no image will be shown in the FlexVision monitor of the exam room. Recall start date: August 1, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips, Allura Xper.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

77920

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 15 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.