Allura systems
Official notice
18 Jul 2025
Last checked
18 Jul 2025
An official Health Canada notice concerns Allura systems. Philips has identified a software issue in the internal communication process between the system software and the x-ray generator firmware which may lead to temporary loss of imaging (x-ray) functionality. This issue may be triggered under either of the following situations: Situation 1 - Pedal tap: When the foot switch is pressed and released rapidly without generating x-rays (due to a short duration of contact), the system may enter a state where x-ray generation is inhibited. When this occurs, the system does not show any message to the user indicating that x-ray is not possible. A warm restart may resolve the situation in some cases. A cold restart will always resolve the situation. This issue can lead to a delay in therapy. Situation 2 - Phase fault: Allura systems are designed to automatically recover from phase faults (a phase fault is an abnormal condition in which one or more phase voltages drop to (near) zero). When a phase fault occurs, the x-ray generator software initiates an automatic recovery process to restore fully system functionality, during which imaging (x-ray) functionality is not available for approximately 5 seconds. When the identified software issue occurs during this recovery, communication with the system software is lost, resulting in the system not recognizing that the generator has recovered and is ready for operation. In this situation, the system shows the user message "generator is busy starting up, no x-ray possible". If the phase fault is detected by the generator during acquisition, the system also shows the message "run aborted: tube problem". To restore system functionality a cold restart of the system is required. These issues affect the Allura Xper systems with software release 8.2 or higher, equipped with a Certeray generator. Recall start date: July 7, 2025 Published 2025-07-18.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77749
Official notice
2025-07-18
The hazard
The hazard
Philips has identified a software issue in the internal communication process between the system software and the x-ray generator firmware which may lead to temporary loss of imaging (x-ray) functionality. This issue may be triggered under either of the following situations: Situation 1 - Pedal tap: When the foot switch is pressed and released rapidly without generating x-rays (due to a short duration of contact), the system may enter a state where x-ray generation is inhibited. When this occurs, the system does not show any message to the user indicating that x-ray is not possible. A warm restart may resolve the situation in some cases. A cold restart will always resolve the situation. This issue can lead to a delay in therapy. Situation 2 - Phase fault: Allura systems are designed to automatically recover from phase faults (a phase fault is an abnormal condition in which one or more phase voltages drop to (near) zero). When a phase fault occurs, the x-ray generator software initiates an automatic recovery process to restore fully system functionality, during which imaging (x-ray) functionality is not available for approximately 5 seconds. When the identified software issue occurs during this recovery, communication with the system software is lost, resulting in the system not recognizing that the generator has recovered and is ready for operation. In this situation, the system shows the user message "generator is busy starting up, no x-ray possible". If the phase fault is detected by the generator during acquisition, the system also shows the message "run aborted: tube problem". To restore system functionality a cold restart of the system is required. These issues affect the Allura Xper systems with software release 8.2 or higher, equipped with a Certeray generator. Recall start date: July 7, 2025
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips, Allura.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 18 Jul 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.