Alinity i total psa reagent kit

Official notice

5 Jun 2025

Last checked

5 Jun 2025

Medium riskHealth Canada

An official Health Canada notice concerns Alinity i total psa reagent kit. An elevated number of complaints have been reported involving both out-of-range third-party quality control (qc) results and a positive bias in patient sample results when using specific product lots. Internal testing with patient samples, including retained lot analysis and trending of patient result medians, confirmed that some lots may exhibit a greater than 10% positive bias. Recall start date: May 22, 2025 Published 2025-06-05.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77561

Official notice

2025-06-05

The hazard

An elevated number of complaints have been reported involving both out-of-range third-party quality control (qc) results and a positive bias in patient sample results when using specific product lots. Internal testing with patient samples, including retained lot analysis and trending of patient result medians, confirmed that some lots may exhibit a greater than 10% positive bias. Recall start date: May 22, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

analysis, trending

Recall Numbers

77561

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Alinity I Total PSA Reagent Kit, Lot or serial number - 73162FZ00 71210FZ00 71210FZ01 71343FZ00 73298FZ00 73155FZ00 71213FZ00 73155FZ01 71346FZ00 73728FZ00, Model or catalog number - 07P9220 07P9230

Risk / hazard

An elevated number of complaints have been reported involving both out-of-range third-party quality control (qc) results and a positive bias in patient sample results when using specific product lots. Internal testing with patient samples, including retained lot analysis and trending of patient result medians, confirmed that some lots may exhibit a greater than 10% positive bias. Recall start date: May 22, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Abbott Ireland Diagnostics Division, Alinity I Total Psa Reagent Kit.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

77561

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 5 Jun 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.