Alcon vitrectomy products

Official notice

20 Aug 2025

Last checked

20 Aug 2025

Medium riskHealth Canada

An official Health Canada notice concerns Alcon vitrectomy products. Alcon is conducting this voluntary medical device field safety correction of impacted standalone vitrectomy consumables, as there is potential for some pouches within impacted lots to have an incomplete seal. While the incomplete seal is visibly noticeable within the pouch, due to the risk that the sterile barrier may be compromised, Alcon is conducting a recall. Should a non-sterile product be used during surgery, post-operative infection may result requiring medical and/or surgical intervention. To date, there have been no reported complaints for this issue from customers for the product in scope. Recall start date: August 11, 2025 Published 2025-08-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77941

Official notice

2025-08-20

The hazard

Alcon is conducting this voluntary medical device field safety correction of impacted standalone vitrectomy consumables, as there is potential for some pouches within impacted lots to have an incomplete seal. While the incomplete seal is visibly noticeable within the pouch, due to the risk that the sterile barrier may be compromised, Alcon is conducting a recall. Should a non-sterile product be used during surgery, post-operative infection may result requiring medical and/or surgical intervention. To date, there have been no reported complaints for this issue from customers for the product in scope. Recall start date: August 11, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77941

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Constellation Vision System Viscous Fluid Control Pak With EnGauge RFID, Lot or serial number - 17DPRT, Model or catalogue number - 8065-7509-57

Product 2

Affected products - Constellation Vision System Viscous Fluid Control Pak With EnGauge RFID, Lot or serial number - 17ERNK, Model or catalogue number - 8065-7509-57

Product 3

Affected products - Chandelier W/RFID, Lot or serial number - 17DPN7, Model or catalogue number - 8065751577

Product 4

Affected products - Chandelier W/RFID, Lot or serial number - 17DPN8, Model or catalogue number - 8065751577

Product 5

Affected products - Endoillumination Probes, Lot or serial number - 17E0TX, Model or catalogue number - 8065750972

Product 6

Affected products - Endoillumination Probes, Lot or serial number - 17C359, Model or catalogue number - 8065751441

Product 7

Affected products - Endoillumination Probes, Lot or serial number - 17C6U5, Model or catalogue number - 8065751441

Risk / hazard

Alcon is conducting this voluntary medical device field safety correction of impacted standalone vitrectomy consumables, as there is potential for some pouches within impacted lots to have an incomplete seal. While the incomplete seal is visibly noticeable within the pouch, due to the risk that the sterile barrier may be compromised, Alcon is conducting a recall. Should a non-sterile product be used during surgery, post-operative infection may result requiring medical and/or surgical intervention. To date, there have been no reported complaints for this issue from customers for the product in scope. Recall start date: August 11, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Alcon Laboratories, Alcon Vitrectomy.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

77941

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Precision1 one-day contact lenses

Medium riskHealth Canada7 Oct 2025
CA
No image published for this recall

Constellation® ultravit® and hypervit® vitrectomy probe

Critical riskHealth Canada26 Sept 2025
CA
No image published for this recall

Constellation® vision system total plus vitrectomy pak

Medium riskHealth Canada30 Jul 2025
CA
No image published for this recall

Total1 and total30 contact lenses

Critical riskHealth Canada3 Sept 2024
CA
No image published for this recall

Alcon ophthalmic knife

Medium riskHealth Canada18 Aug 2024
CA
No image published for this recall

Ngenuity 3d visualization system

Critical riskHealth Canada7 Jun 2024
CA
No image published for this recall

Custom-pak ophthalmic surgery procedure pack

Medium riskHealth Canada2 Apr 2024
CA
No image published for this recall

Custom-pak ophthalmic surgery procedure pack

Medium riskHealth Canada14 Jul 2023
CA
No image published for this recall

Custom-pak™ ophthalmic surgery procedure pack

Medium riskHealth Canada10 Nov 2022
CA
No image published for this recall

Intramedullary bone saw blade assembly

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Glidesheath slender® introducer sheath

Medium riskHealth Canada2 Jul 2026
CA
No image published for this recall

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.