Acuson sequoia diagnostic ultrasound system
Official notice
31 Dec 2021
Last checked
31 Dec 2021
An official Health Canada notice concerns Acuson sequoia diagnostic ultrasound system. System overestimates the circumference when the manual trace function is used. The trace circumference value is overestimated; however, the area value is accurate. The issue is present in Sequoia's VA20A/VA25 onward as the VA20A introduced a new calculation design that did not include an optimization step of discarding pixel points that are too close together, causing the close pixel points into the accumulated circumference/distance value makes the final value highly sensitive to small trackball motions. Recall start date: 2021-12-16 Published 2021-12-31.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63779
Official notice
2021-12-31
The hazard
The hazard
System overestimates the circumference when the manual trace function is used. The trace circumference value is overestimated; however, the area value is accurate. The issue is present in Sequoia's VA20A/VA25 onward as the VA20A introduced a new calculation design that did not include an optimization step of discarding pixel points that are too close together, causing the close pixel points into the accumulated circumference/distance value makes the final value highly sensitive to small trackball motions. Recall start date: 2021-12-16
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Siemens, Acuson Sequoia Diagnostic Ultrasound System.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 31 Dec 2021. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.