No image published for this recall

No image published for this recall

Access 2 reaction vessels

Official notice

13 Nov 2025

Last checked

13 Nov 2025

Medium riskHealth Canada

An official Health Canada notice concerns Access 2 reaction vessels. Instrument testing replicated failures seen in the field (e.g., pipetting, rake, wash carousel errors) for suspect RV lots. Inspections confirmed high percentages of rigid gate vestige exceeding specifications across multiple lots. Published 2025-11-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81215

Official notice

2025-11-13

The hazard

Instrument testing replicated failures seen in the field (e.g., pipetting, rake, wash carousel errors) for suspect RV lots. Inspections confirmed high percentages of rigid gate vestige exceeding specifications across multiple lots.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

81215

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Access 2 Reaction Vessels, Lot or serial number - P243971, Model or catalogue number - 81901

Product 2

Affected products - Access 2 Reaction Vessels, Lot or serial number - P244345, Model or catalogue number - 81901

Product 3

Affected products - Access 2 Reaction Vessels, Lot or serial number - P250005, Model or catalogue number - 81901

Product 4

Affected products - Access 2 Reaction Vessels, Lot or serial number - P250001, Model or catalogue number - 81901

Product 5

Affected products - Access 2 Reaction Vessels, Lot or serial number - P243943, Model or catalogue number - 81901

Product 6

Affected products - Access 2 Reaction Vessels, Lot or serial number - P244202, Model or catalogue number - 81901

Product 7

Affected products - Access 2 Reaction Vessels, Lot or serial number - P244344, Model or catalogue number - 81901

Product 8

Affected products - Access 2 Reaction Vessels, Lot or serial number - P250002, Model or catalogue number - 81901

Product 9

Affected products - Access 2 Reaction Vessels, Lot or serial number - P250004, Model or catalogue number - 81901

Product 10

Affected products - Access 2 Reaction Vessels, Lot or serial number - P250003, Model or catalogue number - 81901

Risk / hazard

Instrument testing replicated failures seen in the field (e.g., pipetting, rake, wash carousel errors) for suspect RV lots. Inspections confirmed high percentages of rigid gate vestige exceeding specifications across multiple lots.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Access 2 Reaction Vessels.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

81215

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Nov 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.