No image published for this recall

No image published for this recall

Access 2 immunoassay analyzer

Official notice

31 Jan 2023

Last checked

31 Jan 2023

Critical riskHealth Canada

An official Health Canada notice concerns Access 2 immunoassay analyzer. Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous results. Recall start date: January 18, 2023 Published 2023-01-31.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72633

Official notice

2023-01-31

The hazard

Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous results. Recall start date: January 18, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72633

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Access 2 Immunoassay Analyzer, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 81600N

Risk / hazard

Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous results. Recall start date: January 18, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Access 2 Immunoassay Analyzer.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

critical

Recall details

Recall number

72633

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 31 Jan 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.