HeartWare Ventricular Assist Device AC adapters and batteries

Official notice

11 Jan 2016

Last checked

11 Jan 2016

High riskACCC

An official ACCC notice concerns HeartWare Ventricular Assist Device AC adapters and batteries. In either case there is the potential for additional risk to the patient from not being connected to back-up power sources and this could lead to serious injury or death if the remaining power source is insufficient or fails. Published 2016-01-11.

What to do now

If you or someone you care for has an implanted HeartWare device check the serial numbers on your AC adapter(s) and batteries. If the serial numbers are within the ranges listed above you should contact your cardiologist or the hospital where the operation was performed to arrange for replacement. If you are unsure of the serial numbers on your AC adapter(s) and batteries, please contact your cardiologist. If an AC adapter fails, the patient's controller will sound an alarm and automatically switch to an alternative power source. The patient or carer should follow HeartWare’s Instructions for Use (IFU) and patient manual and replace the failed AC adapter with a backup AC adapter (or another power source). If batteries drop their charge prematurely an escalating alarm will sound, notifying the patient/carer of a critical battery condition. In such an event, the patient/carer will need to immediately connect an additional power source to the controller to avoid a pump stop. Patients are advised to always have available two fully charged spare batteries and a spare AC adapter. All AC adapters and batteries with serial numbers in the ranges listed above should be replaced. If you have any questions or concerns regarding this issue, please contact your cardiologist.

Recall number

eccf43b0-150c-43d2-8a13-43331cca7eac

Official notice

2016-01-11

The hazard

In either case there is the potential for additional risk to the patient from not being connected to back-up power sources and this could lead to serious injury or death if the remaining power source is insufficient or fails.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

within, are

Models

DESCRIPTION THE SERIAL NUMBERS OF AFFECTED DE

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

The HVAD system takes over some or all of the pumping function of the heart. The device includes a pump which is attached to the patient's heart and is connected via a driveline to an external controller that regulates the pump's function and monitors the system. The system is implanted either in patients awaiting heart transplants or as an alternative to transplantation.

Risk / hazard

In either case there is the potential for additional risk to the patient from not being connected to back-up power sources and this could lead to serious injury or death if the remaining power source is insufficient or fails.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

AU · Australia

Affected area 1

AU · Australia

Source & status

Source

Tier

beta

Category

HOME

Classification

Pharmacy and medical products, Power supply and storage

App estimate

high

Recall details

Recall number

eccf43b0-150c-43d2-8a13-43331cca7eac

Classification

Pharmacy and medical products, Power supply and storage

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 Jan 2016. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.