Enhanced Enlite Sensor (MMT-7008) and Guardian Sensor 3 (MMT- 7020) Glucose Monitoring System
Official notice
21 Mar 2021
Last checked
21 Mar 2021
An official ACCC notice concerns Enhanced Enlite Sensor (MMT-7008) and Guardian Sensor 3 (MMT- 7020) Glucose Monitoring System. If incorrect information is used to make treatment decisions, this could lead to an incorrect dose of insulin. Hydroxurea use results in higher sensor glucose readings compared to blood glucose readings and may result in the following: · Inaccurate graphs or missed alarms and alerts · Delay to or loss of sensor-enabled insulin suspension (sensor-enabled insulin pump use only) · Sensor glucose readings in CareLink reports being substantially higher than blood glucose readings Published 2021-03-21.
What to do now
Medtronic will contact affected users in writing. Medtronic are advising patients: · Do not use continuous glucose monitoring while taking hydroxyurea. · Only use the blood glucose meter to monitor glucose levels while taking hydroxyurea. · Consult with a healthcare professional if users are taking hydroxyurea and using continuous glucose monitoring. · When using an insulin pump system, disable the continuous glucose monitoring feature by going to the sensor settings. Refer to the user guide that came with the insulin pump for instructions. · Always check the label of any medication to confirm with a healthcare professional whether hydroxyurea or hydroxycarbamide is an active ingredient. For further information, contact Medtronic Customer Service by phone on 1800 777 808.
Recall number
e1bf1804-289e-4aa4-aa5b-ef53d6ce26f1
Official notice
2021-03-21
The hazard
The hazard
If incorrect information is used to make treatment decisions, this could lead to an incorrect dose of insulin. Hydroxurea use results in higher sensor glucose readings compared to blood glucose readings and may result in the following: · Inaccurate graphs or missed alarms and alerts · Delay to or loss of sensor-enabled insulin suspension (sensor-enabled insulin pump use only) · Sensor glucose readings in CareLink reports being substantially higher than blood glucose readings
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Source & status
Source & status
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 21 Mar 2021. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.