Astral 100 and Astral 150 ventilators

Official notice

24 Aug 2016

Last checked

24 Aug 2016

Risk not stated by sourceACCC

An official ACCC notice concerns Astral 100 and Astral 150 ventilators. If ventilation stops without prior warning, 'ventilator-dependent' patients who are using the internal battery as the sole power source (for example, during mobile use or when AC power is unavailable or interrupted) may be at risk of serious health problems. Published 2016-08-24.

What to do now

If you or someone you provide care for uses an Astral 100 or Astral 150 ventilator, be aware of this issue. From September 2016, you should be contacted by your local service provider to schedule replacement of the internal battery. You can continue to use your ventilator in the meantime. However, ResMed recommends that the following precautions be taken to minimise risks (especially for ventilator-dependent patients): •Connect the ventilator to AC power whenever possible. In the event of battery failure, connect to AC power immediately to resume ventilation. •For ventilator-dependent patients, use an external power source (Astral External Battery or RPSII battery) including when AC power is unavailable or interrupted. •Use an external power source in mobile situations. Do not rely solely on the internal battery for mobile use. •Ensure the external battery source is sufficiently charged before using in mobile situations. •For ventilator-dependent patients, always have alternate ventilation equipment available, such as a back-up ventilator, manual resuscitator or similar device. •Ventilator-dependent patients should be continuously monitored by qualified personnel or adequately trained carers. These personnel and carers must be capable of taking the necessary corrective action in the event of a ventilator alarm or malfunction. •Return the ventilator to a service provider if it displays any notifications of battery issues or significant degradation in battery performance. If you have any questions or concerns about this issue, speak to the supplier of your device or your local service provider. Alternatively, contact ResMed on 1300 305 705.

Recall number

500e611a-d208-48b2-90e1-b356357b2372

Official notice

2016-08-24

The hazard

If ventilation stops without prior warning, 'ventilator-dependent' patients who are using the internal battery as the sole power source (for example, during mobile use or when AC power is unavailable or interrupted) may be at risk of serious health problems.

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Astral 100 and Astral 150 ventilators provide support to patients with respiratory difficulties resulting from health conditions of varying severity. They can be used in both health facilities and home settings. These ventilators have an internal battery that is designed to deliver continuous power when AC power is disrupted or when a patient is mobile (for example, in a wheelchair). The internal battery is intended to provide power for up to eight hours under normal conditions. When a ventilator is being used with the internal battery as the sole power source, the 'low battery' alarm is designed to activate when there are 20 minutes of ventilation time remaining and the 'critically low battery' alarm is designed to activate when there are 10 minutes of ventilation time remaining.

Risk / hazard

If ventilation stops without prior warning, 'ventilator-dependent' patients who are using the internal battery as the sole power source (for example, during mobile use or when AC power is unavailable or interrupted) may be at risk of serious health problems.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

AU · Australia

Affected area 1

AU · Australia

Source & status

Source

Tier

beta

Category

HOME

Classification

Pharmacy and medical products

App estimate

unspecified

Recall details

Recall number

500e611a-d208-48b2-90e1-b356357b2372

Classification

Pharmacy and medical products

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 24 Aug 2016. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.